Sign In | Join Free | My himfr.com |
|
Ask Lasest Price | |
Brand Name : | Forsentek |
Model Number : | FT01 |
Payment Terms : | T/T,Western Union or Paypal |
Supply Ability : | 2170pcs/month |
Delivery Time : | 3~25 working days |
Flange type reaction torque sensor 1NM 2NM 5Nm 10Nm 20N 50N-m 100Nm 150Nm
Miniature flange type torque transducer FT01 is strain gauge based intended for reaction torque measurement in both clockwise and anticlockwise direction. Miniature torque sensor FT01 offers 1NM / 2NM / 5NM / 10NM / 20N-m / 30NM / 50NM / 100N-m / 150N-m nine rated capacities to choose from,high accracy with maximum 0.3% non-linearity of the full scale.End flanges located on both sides of the torque transducer can be used for bolting between shaft or mounting plates, very compact size makes torque cell FT01 suitable for torque measurement where mounting space is limited. Reaction torque transducer FT01 can be modified or customized as per client's request to meet requirements of different applications.
Dimensions(unit in "mm")
Specifications
Capacity | 1NM / 8.8 lb-in 2NM / 17.7 lbf*in 5NM / 44 lbf-in 10NM / 88 lb*in 20NM / 177 lb-in 30NM / 265 lbf-in 50NM / 442 lbf*in 100NM / 885 lb*in 150NM / 1300 lb-in |
Rated output | 2.0mV/V |
Excitation | 3-15V |
Zero balance | ±0.05mV/V |
Nonlinearity | ±0.3% R.O. |
Hysteresis | ±0.3% R.O. |
Nonrepeatability | ±0.1% R.O. |
Creep(5min) | ±0.1% R.O. |
Safe overload | 150% F.S. |
Ultimate overload | 200% F.S. |
Compensated temp. | -10...+40oC |
Operating temp. | -20...+60oC |
Temp. shift zero | ±0.02% R.O./oC |
Temp. shift span | ±0.02% R.O./oC |
Input resistance | 410±30 ohms |
Output resistance | 350±10 ohms |
Insulation resistance | >2000M ohms |
Ingress protection | IP50 |
Material of element | 0-100Nm @ Aluminum alloy 150Nm @ Alloy steel |
Cable | Ø3*2000mm 4-core shielded |
Wiring code | Red--Excitation+ Black--Excitation- Green--Signal+ White--Signal- |
Torque sensor: A transducer that converts mechanical torque into mV
signal proportional to the applied torque
Wheatstone Bridge:A Wheatstone bridge is a network of four
resistive legs. One or more of these legs can be active sensing
elements. The Wheatstone bridge is the electrical equivalent of two
parallel voltage divider circuits. R1 and R2 compose one voltage
divider circuit, and R4 and R3 compose the second voltage divider
circuit. The output of a Wheatstone bridge is measured between the
middle nodes of the two voltage dividers.
Excitation: The voltage applied to the input terminals of the load
cell.
Rated Capacity: The maximum axial load the load cell is designed to
measure within its specifications.
Non-linearity: The maximum deviation of the calibration curve from
a straight line drawn between the no-load and rated outputs;
expressed as a percentage of the rated output and measured on
increasing load only.
Non-repeatability: The maximum difference between load cell output
readings for repeated loadings under identical loading and
environmental conditions.
Hysteresis: The maximum difference between load cell output
readings for the same applied load; one reading obtained by
increasing the load from zero and the other by decreasing the load
from rated output; usually measured at half rated output and
expressed in percent of rated output.
Creep: The change in load cell output occurring with specified
period of time, while under load, and with all environmental
conditions and other variables remaining constant; usually measured
with rated load applied.
Safe Overload: The maximum load in percent of rated capacity which
can be applied without producing a permanent shift in performance
characteristics beyond those specified.
Brief instruction of Rated output(or Sensitivity):
Definition: The signal produced by load cell(or force transducer)
when rated load is applied, expressed in terms of mV/V. The rated
output indicates the output signal range of a sensor.
Sensor output = (Power supply * Rated output) * Load / Capacity
Rated output is normally within 1mV/V~3mV/V.
Power supply is normally within 3-15V.
Load / Capacity is within 0-1.
So the load cell output signal is normally within
0-45mV(-45mV...+45mV for bidirectional load cell).
Example: 2.0mV/V torque sensor with 10V power supply,the output
range is -20mV...+20mV.
![]() |